Fresenius Medical Care 2008T Troubleshooting Manual page 505

Hemodialysis system
Hide thumbs Also See for 2008T:
Table of Contents

Advertisement

Yes
-250 mmHg holds! A)
segments then; B) Leaving [Dialysate Flow] off, proceed to
TMP-
4.0.0.
No
-250 could not be achieved or it DID NOT hold! Either the transducer protector is wet OR a
Four-Way assembly tubing connection is leaking! See parts A and B below:
A) Locate and repair the leak. Consider using another Four-Way assembly!
B) Return to (ABOVE) procedure number
IMPORTANT!
Remove the clamps from both Four-Way tubing
LEFT BLANK INTENTIONALLY
2008T Troubleshooting Guide
P/N 490292 Rev. A
page 509
, procedure number
TMP- 2.0.16
(page 499).
Page 501