Both Pumps Drawing Concentrate/Isolate Leaks - Fresenius Medical Care 2008T Troubleshooting Manual

Hemodialysis system
Hide thumbs Also See for 2008T:
Table of Contents

Advertisement

CO- 1.0.3.8 BOTH PUMPS DRAWING CONCENTRATE / ISOLATE LEAKS
A) ENSURE FIRM connections to the acid and bicarb jugs.
B) Figure below, ENSURE no leaks from the ACID and BICARB pumps.
C) Figure below, if a 'Quick Connector' is present, at the end of the drain tubing, an ADAPTOR is required!
D) As seen in the Figure right, point the drain tubing
OPENING UP at 45° and no higher than two (2) feet
IF POINTED DOWN GRAVITY
above the floor!
FLOW RESULTS IN ERROR!
E) Watch for ONE (1) FULL MINUTE THEN consider BOTH
scenarios next page:
2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 351