Removal and Installation Procedures
Front Cushion
6- 24
Figure 6- 13: Line cord and cabinet removal
1. Assemble equipment and locate modules to be removed:
a. You will need a #2 Phillips screwdriver (Items 1 and 3).
b. Locate the front cushion in the External modules diagram, Figure 6- -2.
2. Orient the analyzer: Set the analyzer so its bottom is down on the work
surface and its front is facing you.
3. Remove the front cushion: See Figure 6- -14.
a. Remove the two screws securing the front cushion to the side panel at
each side.
b. Lifting the front of the analyzer slightly, move the front cushion away.
4. Reinstallation: Reverse step 3 to reinstall the front cushion.
RSA3303A & RSA3308A Service Manual