E4990A Help
WriteData(i * 2 - 1) = ActiveSheet.Cells(i + 10, 3).Value
Next i
DataType = Cells(5, 2)
Select Case DataType
Case "Ascii"
Analyzer.WriteString ":FORM:DATA ASC", True
Analyzer.WriteString "CALC1:DATA:FDAT ", False
Analyzer.WriteList WriteData, ASCIIType_R8, ",", True
Case "Binary"
Analyzer.WriteString ":FORM:DATA REAL", True
Analyzer.WriteIEEEBlock ":CALC1:DATA:FDAT ", WriteData, True
End Select
'*** End procedure
Analyzer.IO.Close
End Sub

Processing Measurement Results

Processing Measurement Results
Searching Peak
Setting the Limit Test Functions
Searching Peak
Overview
Sample Program in Excel VBA
Other topics about Processing Measurement Results
Overview
The program listed in the below section is written in VISA-COM with Excel
VBA. It can be executed from the external PC controller. The program
demonstrates how to search peak.
Searching Peak Sample Program in Excel VBA
Dim ioMgr As VisaComLib.ResourceManager
Dim Ana As VisaComLib.FormattedIO488