2. Fasten two of the four BNC connectors to the corresponding test
connectors by gradually turning the BNC connectors' rotation levers
until each pair of connectors is securely connected. Be sure to align
the grooves on both sides.
3. Simultaneously turn the fixture's two fastening screws clockwise so
that the fixture is secured to the instrument.
4. Finally, secure the remaining two BNC connectors of the fixture by
turning their rotation levers clockwise.
16047E Test Fixture.
Turn ON the Power
Press the power switch to turn ON the power to the E4990A. The E4990A
performs a power-on self-test. During the self-test, the model name,
firmware revision number/date, options, copyright notice, and other
information appear on the LCD. When the self-test is completed, the
measurement screen appears on the LCD.
Specifying Measurement Conditions
Initialize the E4990A to the Preset State
•
Select Fixture
•
Select |Z|-Theta as the Measurement Parameter
•
Select Frequency in Log as the Sweep Parameter
•
Set the Sweep Start/Stop Value
•
Set the Measurement Speed
•
Other topics about Measurement Example
Next, you need to specify how your E4990A should perform the
measurement.
Reverse the above procedure when removing the