Specifying Measurement Conditions - Keysight E4990A Help Manual

Impedance analyzer
Hide thumbs Also See for E4990A:
Table of Contents

Advertisement

2. Fasten two of the four BNC connectors to the corresponding test
connectors by gradually turning the BNC connectors' rotation levers
until each pair of connectors is securely connected. Be sure to align
the grooves on both sides.
3. Simultaneously turn the fixture's two fastening screws clockwise so
that the fixture is secured to the instrument.
4. Finally, secure the remaining two BNC connectors of the fixture by
turning their rotation levers clockwise.
16047E Test Fixture.
Turn ON the Power
Press the power switch to turn ON the power to the E4990A. The E4990A
performs a power-on self-test. During the self-test, the model name,
firmware revision number/date, options, copyright notice, and other
information appear on the LCD. When the self-test is completed, the
measurement screen appears on the LCD.

Specifying Measurement Conditions

Initialize the E4990A to the Preset State
Select Fixture
Select |Z|-Theta as the Measurement Parameter
Select Frequency in Log as the Sweep Parameter
Set the Sweep Start/Stop Value
Set the Measurement Speed
Other topics about Measurement Example
Next, you need to specify how your E4990A should perform the
measurement.
Reverse the above procedure when removing the
Quick Start
43