Reading the marker position time or relaxation time

You can use the following procedure to select how the sweep parameter
value is displayed in the Marker Sweep Parameter Value field: time relative
to the sweep start or relaxation time. This function allows you to read the
sweep parameter value in a unit of time at each point on the trace in the
zero span sweep.
1. Press the
2. Press X Unit > Off|Time|1/2πf.
3. Select the display method of the marker sweep parameter value.
Parameter
Off
Time
1/2πf
sweep parameter value (marker X-axis) only when the frequency is
selected as the sweep parameter. When you use the ∆ marker, the
relative value from the ∆ marker position is displayed in the Marker
Sweep Parameter Value area, regardless of the selection of the
marker sweep parameter value (marker X-axis) display.
Analyzing the Equivalent Circuit and Simulate the Frequency
Characteristics
Calculate the Equivalent Circuit Parameter based on the
Measurement Result
Simulate the Frequency Characteristics based on the Equivalent
Circuit Parameter
Other topics about Analysis and Processing of Result
Calculate the Equivalent Circuit Parameter based on the Measurement Result
Calculate the equivalent circuit parameter based on the measurement
result trace according to the followings steps:
key to display the Marker Function menu.
Marker Function
Definition
Value of the parameter selected as the sweep
parameter (frequency, oscillator level, or dc bias
level)
Time between the sweep start and the completion
of the measurement at the marker position
Relaxation time = 1/2πf (f:measurement
frequency)
You can select 1/2πf as the method to display the marker
Measurement
163