42941A 3.5Mm Port Extension - Keysight E4990A Help Manual

Impedance analyzer
Hide thumbs Also See for E4990A:
Table of Contents

Advertisement

E4990A Help
2. When performing calibration at the test port of the 42941A with a
user-defined calibration kit, perform user calibration with the proper
calibration kit. Refer to
procedure. User calibration is not required when you do not use a
user-defined calibration kit.
3. Connect a probe adapter to the test port of the 42941A Impedance
Probe. See the operation manual of the 42941A for the appropriate
connection setup.
4. Perform the fixture compensation at the device contacts (electrodes)
of the probe adapter. See
compensation procedure.
of a 42941A that is also connected to the four-terminal pair port
of the E4990A, port extension compensation is not required.
Other topics about Calibration
G. 42941A 3.5mm Port Extension
Perform the following calibration when the 3.5-mm port of a 42941A
connected to the E4990A is extended with a one-port configuration and the
device is connected to the cable's end with a minimal connection.
114
User Calibration
Fixture Compensation
When a probe adapter is connected to the test port
for the user calibration
for the fixture