Adapter Selection; Adapter Setup - Keysight E4990A Help Manual

Impedance analyzer
Hide thumbs Also See for E4990A:
Table of Contents

Advertisement

*1. Phase compensation.
*2. Although it has a cable, it is considered a direct-connection-type test
fixture.

Adapter Selection

The adapter selection setting must be confirmed before measurement in
accordance with the connection status of each accessory. If necessary, it
must be changed to the correct setting.
1. Press
Cal
2. Select the corresponding adapter selection from the following:
Adapter Selection
No Adapter
Four-terminal pair, 1 meter
Four-terminal pair, 2 meter
42942A Terminal Adapter
42942A Terminal Adapter and extension cable
42941A Impedance Probe
42941A Impedance Probe and extension cable
through the display of the instrument status area.
the E4990A is equipped with option 120.

Adapter Setup

Adapter Setup
The adapter setup is a process used to acquire the setup data of the
adapter selected by the
performed when using the following accessories with the E4990A for the
first time after it is delivered. The adapter setup should be performed again
when the accessory is changed. Even if the different accessory with the
> Adapter.
Adapter selection status can also be confirmed
The 42941A and 42942A can be selected only when
Adapter
Selection. The adapter setup is normally
Measurement
Softkey Label
NONE
4TP 1M
4TP 2M
7mm 42942A
42942A Extension
Probe 42941A
42941A Extension
63