Error messages during the device check

Message
No communication control- /
charge-board
System leakage
No test lung
Breathing valve inop
Pressure measurement inop
PEEP-valve inop
Patient flow measurement
inop
Instructions for Use Oxylog 2000 plus SW 1.n
Cause
Device defective
Leak in ventilation hose system
and/or test lung.
Test lung not connected or mayor
leakage
PEEP >10 mbar has been set but
not confirmed.
The ventilation hose system has
not been connected correctly.
Pressure measurement is not pos-
sible.
Internal leak in system
Device defective
Flow measurement implausible
Problem Solving
Explanation/Remedy
Contact your local DrägerService
for additional support.
Check hoses, breathing valve,
flow sensor and test lung for leaks
and replace if necessary.
Connect test lung.
Check hoses, breathing valve,
flow sensor and test lung for leaks
and replace if necessary.
The required setting of
PEEP >10 mbar is only possible
when confirmed via the rotary
knob.
Connect ventilation system cor-
rectly.
Contact your local DrägerService
for additional support.
Check hoses, breathing valve,
flow sensor and test lung for leaks
and replace if necessary.
Contact your local DrägerService
for additional support.
Replace flow sensor.
Contact your local DrägerService
for additional support.