System Overview
Reusable hose system
A
E
D
A Breathing valve
B Ventilation hose
C Flow and pressure measuring hoses
D Angled connector
E Flow sensor
Available ventilation modes
– VC-CMV / VC-AC
Volume Controlled - Controlled Mechanical
Ventilation with PEEP.
Volume Controlled - Assist Control with PEEP.
– VC-SIMV (PS)
Volume Controlled - Synchronized Intermittent
Mandatory Ventilation with PEEP
(Optionally with Pressure Support).
Procedure for weaning patients off the ventilator
after they have started spontaneous breathing.
14
B
C
Disposable hose system
D
E
A Breathing valve
B Ventilation hose
C Flow and pressure measuring hoses
D Flow sensor
E Angled connector
– SpnCPAP (PS)
Continuous Positive Airway Pressure
(Optionally with Pressure Support)
Spontaneous breathing with positive airway
pressure.
Instructions for Use Oxylog 2000 plus SW 1.n
A
B
C