Alarms
Setting alarm limits automatically
WARNING
After using the function Auto alarm limits:
check if the new alarm limits are appropriate
for the patient.
Risk of hypoventilation.
The function Auto alarm limits sets the alarm
limits on the basis of the following actual measured
values at the time of activation:
MV
: Measured value MV +2 L/min
MV
: Measured value MV –2 L/min
RRsp
: Measured value RRsp +5/min, with a
minimum of 10/min.
This automatic selection of alarm limits is perfor-
med only once, when confirmed, via the rotary
knob. The alarm limits refer to the current measu-
red values for MV and RRsp.
64
Instructions for Use Oxylog 2000 plus SW 1.n