Assembly
O xylog 2000 plus
C
B

Connecting the gas supply

Take care when handling O
WARNING
Secure O
cylinders so they cannot fall over
2
and keep away from excessive heat.
Risk of explosion!
WARNING
Do not grease or lubricate O
cylinder valves and pressure reducers and do
not handle with greasy hands.
Risk of fire!
WARNING
Operate cylinder valves by hand and rotate
smoothly to prevent the risk of fire or
explosion.
Do not use tools.
WARNING
No smoking or open flames.
O
is combustible and can intensity fires.
2
34
A
2:
fittings, such as
2
1 Connect the mains plug (A) to the mains outlet.
2 Connect the DC plug (B) to the DC outlet on the
Oxylog 2000 plus.
3 When the Oxylog 2000 plus is connected to an
external supply, the indicator
and displays the internal battery status.
WARNING
Only use medical grade oxygen that is dry and
free from dust and oil.
Contaminated gas can cause device
malfunction.
WARNING
Always provide adequate ventilation in order
to maintain ambient O
to prevent risk of fire.
WARNING
Always use extreme caution when using oxy-
gen, to prevent risk of fire.
Instructions for Use Oxylog 2000 plus SW 1.n
N
(C) lights up
concentration < 24%,
2