Pressure Support (Optional) - Dräger Oxylog 2000 plu Instructions For Use Manual

Emergency and transport ventilator software 1.n
Table of Contents

Advertisement

The parameters RRapn and VTapn, which are
required for setting apnea ventilation, are now
displayed:
3 Set RRapn and VTapn.
4 Set Pmax. This determines the maximum air-
way pressure allowed during apnea ventilation.
The ventilation time ratio I:E = 1:1.5 and the plateau
time Tplat % = 0 are preset during apnea
ventilation.
To switch apnea ventilation OFF
Set Tapn to OFF.
To end apnea ventilation
Press the Alarm Reset key.
The ventilator resumes ventilating with the original
mode and parameter settings.
The manufacturer default settings are:
– RRapn = 12 /min
– VTapn = 500 mL
These default settings can be configured.
Refer to the "Customer Service Mode" section for
additional information.
NOTE
Apnea ventilation can only be activated in
the ventilation mode SpnCPAP without NIV.
The minimum ventilation required by the patient
must be monitored via the lower alarm limit
MV
.
Instructions for Use Oxylog 2000 plus SW 1.n

Pressure support (optional)

The following can additionally be set on the display
for SpnCPAP / PS:
– Sensitivity Trigger (for synchronization with the
patient's spontaneous breathing efforts).
Successful patient triggering is briefly indicated
by an asterisk (*) in the middle of the status
alarm messages window.
– Pressure support
– Pressure rise time Slope (for pressure
support
Psupp).
Operation
Psupp above PEEP.
53