Intended Use - Dräger Oxylog 2000 plu Instructions For Use Manual

Emergency and transport ventilator software 1.n
Table of Contents

Advertisement

Application

Intended use

®
The Oxylog
2000 plus is a time-cycled, volume
controlled emergency and transport ventilator with
pressure support for patients requiring mandatory
or assisted ventilation with a tidal volume of 100 mL
upwards.
For use by and under the supervision of trained
health care professionals.
Indications/Contraindications
For patients requiring a tidal volume of 100 mL
upwards.
WARNING
The Oxylog 2000 plus ventilator may only be
used under the supervision of qualified medi-
cal personnel in order to be able to provide
immediate corrective action in case of a mal-
function.
Environment of use
For use in the following environments:
– Mobile use for emergency patients, for outdoor
and indoor environments.
– During transport in ambulances or aircraft,
including helicopters.
– In accident and emergency departments.
– When moving ventilated patients around the
hospital.
– In the recovery room.
10
WARNING
Do not use the equipment in hyperbaric
chambers!
The device may malfunction, causing danger
to the patient.
WARNING
Do not use the equipment in conjunction with
magnetic resonance imaging (MRI, NMR, NMI).
The device may malfunction, causing danger
to the patient.