Getting Started; Device Check - Dräger Primus Infinity Empowered Quick Reference Manual

Hide thumbs Also See for Primus Infinity Empowered:
Table of Contents

Advertisement

Getting Started

NOTE
The Primus Infinity Empowered Quick Reference Guide is not a replacement or substitute for the Instructions for
Use. Any use of the device requires full understanding and strict observation of the Instructions for Use.

Device check

Power on
1
Switch on Primus Infinity Empowered: press the
system power switch
sounds.
After approx. 15 seconds all LEDs and the loud-
speaker are tested by Primus Infinity Empowered.
The initial screen appears after about 20 seconds. Pri-
mus Infinity Empowered now loads its software and
tests its internal memory.
Quick Reference Guide Primus Infinity Empowered SW 4.5n
Check List
The Check List is displayed after about 35 seconds.
A
The test steps to be performed are grouped in five cat-
egories:
(A), an acoustic tone
A Gas Supply
B Vaporizers
C Breathing Circuit
D Miscellaneous
Primus
Check List
B
A
C
D
pipeline pressure
cylinder pressure
O
flush functional?
2
Safety O
control functional?
2
Correctly locked in positon?
Set to zero?
Fill level OK?
Safety filler locked?
Fully assembled?
Correctly connected?
Gas scavenger connected and flow adjust-
ed?
Soda lime OK?
Water trap fill level OK?
Suction OK?
Emergency resuscitator present and func-
tional?
Getting Started
F
G
E
H
17