Operation
SpnCPAP (optional PS)
Continuous Positive Airway Pressure
WARNING
Only use SpnCPAP for patients with sufficient
spontaneous breathing.
Otherwise there is a risk of the patient
receiving insufficient ventilation.
Spontaneous breathing can optionally be assisted
with PS and NIV.
Set the ventilation pattern with the controls below
the display:
– Maximum airway pressure Pmax.
– O
setting FiO
.
2
2
Apnea ventilation
Paw
Pmax
Apnea alarm time
Tapn
Start
Flow
apnea ventilation
with apnea alarm
52
Apnea back-up ventilation is only applicable when
using the SpnCPAP mode. In the event of an
apnea, the ventilator will automatically activate
volume-controlled mandatory ventilation
(VC-CMV).
When an apnea occurs, the device simultaneously
issues an alarm signal and switches to volume con-
trolled ventilation with the parameters respiratory
rate RRapn, tidal volume VTapn, and the maxi-
mum airway pressure Pmax when the apnea time
Tapn has been reached. The ventilation time ratio
I:E is set to 1:1.5. The plateau time Tplat % is 0.
The patient can breathe spontaneously during
apnea ventilation. The mandatory frequency
RRapn remains constant.
Setting apnea ventilation
t
1
RR
On the display:
1 Press the Settings
t
2 Set Tapn with the rotary knob (B) to a value
C
2/3
page
appears.
between 15 and 60 seconds.
Instructions for Use Oxylog 2000 plus SW 1.n
Oxylog 2000 plus
A
B
key (A) until