Spncpap (Optional Ps); Apnea Ventilation - Dräger Oxylog 2000 plu Instructions For Use Manual

Emergency and transport ventilator software 1.n
Table of Contents

Advertisement

Operation

SpnCPAP (optional PS)

Continuous Positive Airway Pressure
WARNING
Only use SpnCPAP for patients with sufficient
spontaneous breathing.
Otherwise there is a risk of the patient
receiving insufficient ventilation.
Spontaneous breathing can optionally be assisted
with PS and NIV.
Set the ventilation pattern with the controls below
the display:
– Maximum airway pressure Pmax.
– O
setting FiO
.
2
2

Apnea ventilation

Paw
Pmax
Apnea alarm time
Tapn
Start
Flow
apnea ventilation
with apnea alarm
52
Apnea back-up ventilation is only applicable when
using the SpnCPAP mode. In the event of an
apnea, the ventilator will automatically activate
volume-controlled mandatory ventilation
(VC-CMV).
When an apnea occurs, the device simultaneously
issues an alarm signal and switches to volume con-
trolled ventilation with the parameters respiratory
rate RRapn, tidal volume VTapn, and the maxi-
mum airway pressure Pmax when the apnea time
Tapn has been reached. The ventilation time ratio
I:E is set to 1:1.5. The plateau time Tplat % is 0.
The patient can breathe spontaneously during
apnea ventilation. The mandatory frequency
RRapn remains constant.
Setting apnea ventilation
t
1
RR
On the display:
1 Press the Settings
t
2 Set Tapn with the rotary knob (B) to a value
C
2/3
page
appears.
between 15 and 60 seconds.
Instructions for Use Oxylog 2000 plus SW 1.n
Oxylog 2000 plus
A
B
key (A) until