When Using A Bacterial Filter Or Hme; Hose Connections - Dräger Oxylog 2000 plu Instructions For Use Manual

Emergency and transport ventilator software 1.n
Table of Contents

Advertisement

Assembly

When using a bacterial filter or HME

NOTE
When using a bacterial filter or HME, measured
flows may deviate from the expiratory flows, as
temperature and humidity of the gas are reduced.
Connect the bacterial filter or HME to the angled
connector.
WARNING
Bacterial filters increase the exhalation resis-
tance and dead space volume of the ventila-
tion system.

Hose connections

B
B
1 Connect the ventilation hose (A) to the
breathing valve.
2 Connect the flow measuring hoses (B) to the
nozzles on the flow sensor. Note the different
diameters.
30
A
CAUTION
Do not use electrically conductive hoses!
Risk of electric shock.
This can endanger the patient.
3 Connect the flow measuring (A) hoses to the
Oxylog 2000 plus.
4 Connect the ventilation hose (B) to the gas out-
put on the Oxylog 2000 plus.
Instructions for Use Oxylog 2000 plus SW 1.n
A
B