Preparing ventilation mode

To activate the ventilation mode

1 Press and hold the ventilation mode key for
approximately 3 seconds.
Or
2 Press the ventilation mode key and confirm by
pressing the rotary knob.
The new ventilation mode selected is now effective.

Set ventilation parameters

1 Set the required control below the display.
Or
2 Select, set and confirm a parameter on the
display with the rotary knob.
The former settings are retained if confirmation is
not received within 15 seconds. Attention is drawn
to this fact by the advisory message ! Settings not
confirmed.
When the PEEP-setting is increased above 10
mbar, a message Confirm PEEP above 10 mbar?
will appear to request confirmation of the change.
The PEEP setting can be increased to the desired
testting after the message is acknowledged with
the rotary knob.
Instructions for Use Oxylog 2000 plus SW 1.n