Connect the disposable hose system

WARNING
Do not use disposable hose systems other
than those on the "List of Accessories". The
minute volume may be measured incorrectly
and the device may malfunction.
A
WARNING
Ensure that the flow measuring hoses are cor-
rectly positioned, otherwise the volume will be
measured incorrectly.
1 Connect the blue flow measuring hose (B) to
the blue gas outlet.
2 Connect the transparent flow measuring hose
(A) to the other gas outlet.
3 Connect the ventilation hose (C) to the gas out-
let on the Oxylog 2000 plus.
Instructions for Use Oxylog 2000 plus SW 1.n
B
C

When using a bacterial filter or HME

NOTE
When using a bacterial filter or HME, measured
flows may deviate from the expiratory flows, as
temperature and humidity of the gas are reduced.
Connect the bacterial filter or HME.
WARNING
Bacterial filters increase the exhalation resis-
tance and dead space volume of the ventila-
tion equipment.
When changing the ventilation hose
system
If the reusable ventilation hose system is to be used
instead of a disposable hose system or vice versa:
1 Have the nozzles on the device changed by
trained service specialists.
2 Reconfigure the device accordingly. Refer to
the "Customer Service Mode" on page 72 for
additional information.