Cleaning, Disinfection and Sterilization
Disinfecting by wiping
Ventilator and medical gas hose:
– Follow the manufacturer's instructions. Remove
heavy soiling with a disposable cloth first.
WARNING
Do not allow any liquid to enter the ventilator
or medical gas hose!
Risk of malfunction.
92
Bath disinfection for reusable hose system
Disassembled parts of the breathing valve, flow
sensor, ventilation hose and flow measuring hoses:
WARNING
Follow the manufacturer's instructions.
Agitate parts thoroughly in the solution.
Do not clean with a hard brush!
WARNING
Do not allow any objects to enter the breathing
valve or flow sensor!
Risk of malfunction.
WARNING
Rinse parts thoroughly with distilled water.
Disinfectant residues can cause the rubber
disc to become jammed in the breathing
valve!
WARNING
Allow to dry completely. The breathing valve
and flow measuring hoses may not function
correctly if water remains in these parts!

Sterilizing the reusable hose system

– Disassemble the breathing valve, flow sensor
and angled connector. When disassembling the
breathing valve from the flow sensor, pull in a
straight line.
Do not rotate the parts, as this may damage the
flow sensor. Dismantle the breathing valve.
– The disassembled parts of the breathing valve,
the flow sensor, the angled connector, the flow
measuring hoses and the ventilation hose can
be sterilized in hot steam at 134
dance with EN 285 (Sterilization – Steam
sterilization – Large-scale sterilization) for at
least 3 minutes, up to 10 minutes.
– The hose system can be sterilized up to a
100 times.
Instructions for Use Oxylog 2000 plus SW 1.n
o
C in accor-