Dräger Oxylog 2000 plu Instructions For Use Manual page 115

Emergency and transport ventilator software 1.n
Table of Contents

Advertisement

Immunity against
surges on AC mains
lines
(IEC 61000-4-5)
power frequency
magnetic field
50/60 Hz
(IEC 61000-4-8)
voltage dips and
short
interruptions on AC
mains input lines
(IEC 61000-4-11)
radiated RF
(IEC 61000-4-3)
RF coupled into lines
(IEC 61000-4-6)
I
Information regarding separation distances (IEC 60601-1-2, tables 205 and 206).
X1)
For P
the highest possible "equivalent isotropic radiated power" of the adjacent RF transmitter has to be inserted
EIRP
(value in Watt). Also in the vicinity of equipment marked with the symbol
fixed, portable or mobile RF transmitters at the location of the Oxylog 2000 plus should be less than 3 V/m in the frequency
range from 150 kHz to 2.5 GHz and less than 1 V/m above 2.5 GHz.
X2
)
ISM bands in this frequency range are: 6.765 MHz - 6.795 MHz, 13.553 MHz - 13.567 MHz,
26.957 MHz - 27.283 MHz, 40.66 MHz - 40.70 MHz.
Instructions for Use Oxylog 2000 plus SW 1.n
IEC 60601-1-2 test level
common mode:
2 kV
differential mode: 1 kV
3 A/m
dip >95%, 0.5 periods
dip 60%, 5 periods
dip 30%, 25 periods
dip >95%, 5 seconds
80 MHz - 2,5 GHz: 10 V/m
150 kHz - 80 MHz: 10 V
within ISM bands,
3 V outside ISM bands
Compliance
level (of the
Oxylog 2000
plus)
2 kV
1 kV
N/A
N/A
30 V/m
10 V
(X2)
10 V
interference may occur. Field strengths from
Technical Data
Electromagnetic environment
Mains power quality should be
that of a typical commercial or
hospital environment.
Test is not necessary because
the Oxylog 2000 plus is not likely
to be sensitive to magnetic field
disturbances, such as CRT
monitors or hall elements.
No voltage dips or short interrup-
tions, since a fully charged bat-
tery must be installed for safety
reasons, even when operating
from an external supply.
Recommended separation dis-
tance from portable and mobile
RF transmitters with transmis-
sion power P
to the
EIRP
Oxylog 2000 plus including its
lines:
EIRP (X1)
1.84 m *
P
Recommended separation dis-
tance from portable and mobile
RF transmitters with transmis-
sion power P
to the
EIRP
Oxylog 2000 plus including its
lines:
1.84 m *
P
EIRP
(X1)
115