Remove The Disposable Hose System - Dräger Oxylog 2000 plu Instructions For Use Manual

Emergency and transport ventilator software 1.n
Table of Contents

Advertisement

Cleaning, Disinfection and Sterilization
WARNING
The rubber disc (A) in the housing must not be
removed, damaged or bent, otherwise the
valve will not work properly and endangers
the patient.
Risk of CO
rebreathing.
2

Remove the disposable hose system

A
1 Disconnect flow measuring hoses (A and B).
2 Disconnect the ventilation hose (C).
3 Correctly dispose of the complete disposable
hose system. Refer to the chapter "Disposal"
on page 99.
CAUTION
The disposable hose system must not be
sterilized: it cannot withstand high temperatures
and may be damaged!
90
B
C
Instructions for Use Oxylog 2000 plus SW 1.n