Device quick check

Set language for display messages –
see page 39
Device quick check Oxylog 2000
Before carrying out this device quick check, first check
readiness for operation as described on page 28.
The device check must be carried out before each use.
Any operation of the device requires thorough knowledge
of the Instructions for Use.
Type:
Serial No.:
Check before start-up:
cylinder pressure equals at least 100 bar or
correct connection to a piped medical gas supply
Power supply assured:
fully charged pack of rechargeable or new batteries
If external power supply:
either connected to mains
or to on-board power supply
All tubing connected
Function check
Connect test lung to breathing valve
Rotary knobs: V
to 0.5 L
T
Freq. to 12 1/min
T
: T
to 1: 2
I
E
Pmax to 60 mbar
PEEP to 5 mbar
Switch for ventilation modes set to IPPV
ON/OFF switch to I:
Display:
Self test
SW-version xx.xx
Red alarm indicator lights up briefly
Alarm tone sounds twice
Green LED lights up with external power supply
After approx. 6 seconds
Display:
Self test OK
6
Oxylog 2000 ventilates the test lung
Test Paw low alarm:
Remove test lung, alarm after approx. 20 seconds
Test Paw high alarm:
Keep test lung compressed. Alarm
For fault messages, see Instructions for Use
"Fault, Cause, Remedy", page 34.
Device check completed