Error messages during the device check

Message in the
alarm window
No communication
control- / charge-board
System leakage
No test lung
Breathing valve inop
Pressure
measurement inop
PEEP-valve inop
Cause
Device defective.
Leak in ventilation hose system and/or test lung.
Test lung not connected or major leakage.
Breathing valve has malfunctioned.
The ventilation hose system has not been
connected correctly.
Pressure measurement is not possible.
Internal leak in system.
Device defective.
Explanation/Remedy
Contact your local DrägerService
Check hoses, breathing valve,
flow sensor and test lung for leaks and
replace if necessary.
Connect test lung.
Check hoses, breathing valve, flow sensor
and test lung for leaks and replace if necessary.
Check correct condition of breathing valve
including diaphragm and rubber disc; fit a new
breathing valve if necessary or use a new
disposable hose set.
Connect ventilation system correctly.
Contact your local DrägerService for additional
support.
Check hoses, breathing valve, flow sensor and
test lung for leaks and replace if necessary.
Contact your local DrägerService for additional
support.
49
Trouble Shooting