Gas Supply - Dräger Oxylog 2000 plu Instructions For Use Manual

Emergency and transport ventilator software 1.n
Table of Contents

Advertisement

Technical Data
AC/DC power pack
DC/DC converter

Gas supply

1) NIST = Non Interchangeable Screw Thread Connection
2) DISS = Diameter Index Safety Systems
3) N-F = French standard
110
Temperature range
Protection class to
EN 60601
Input
Output
Temperature range
Input
Output
O
service pressure
2
Supply gas
O
inlet connection
2
Gas cylinders and
pressure reducers
Pressure reducer
Gas consumption for
internal control
Accuracy of gas
consumption indication
o
o
–20
C to 50
C
Class II, the earthing is used for EMC purposes.
100 to 240 V~/ 50 to 60 Hz / 0.6 - 0.3 A
19 V ±0.5 V / 2.1 A
o
o
–20
C to 50
C
12 / 24 / 28 V DC / 9 A
19 V ±0.5 V / 2.1 A
From a pipeline system or from a medical gas
cylinder.
270 kPa to 600 kPa maximum 100 L/min
Medical oxygen.
either:
1)
NIST
to EN 739, or
2)
DISS
to CGA V5-1989, or
3)
N-F
S90-116
WARNING
Only use medical grade oxygen that is dry and
free from dust and oil.
Contaminated gas can cause device malfunction.
WARNING
Only use compressed gas cylinders and pres-
sure reducers, which comply with all applicable
regulations and have been approved.
Must have a vent valve on the output side to limit the
delivery pressure to approximately 1000 kPa in the
event of a fault.
Average 0.5 L/min
15% or ±1 L/min, whichever is greater.
Instructions for Use Oxylog 2000 plus SW 1.n