Starting Operation; Switch On - Dräger Oxylog 2000 plu Instructions For Use Manual

Emergency and transport ventilator software 1.n
Table of Contents

Advertisement

Operation

Starting operation

The actual screen display may differ in appearance
or configuration.
WARNING
Only use a ventilator that has been cleaned
and successfully tested for operation, to pre-
vent a health risk for the patient and user.

Switch ON

Briefly press the
O
The Oxylog 2000 plus performs a self-test.
– The sef-test will be completed in approxi-
mately six seconds.
During the self-test, the system briefly displays the
starting page with the software version and a
prompt for the operator to select the configuration
menu, or to activate the device check by pressing
the rotary knob.
The bar graph indicates the progress of the self-
test.
46
Oxylog 2000 plus
A
key (A).
Upon completion of the self-test, the ventilator
automatically begins ventilation with the default
settings.
The opening display with configured settings is
displayed if the central rotary knob is not pressed.
The manufacturer's default settings are:
– Ventilation mode VC-CMV.
– Ventilation time ratio I:E = 1:1.5.
– Positive end expiratory pressure
PEEP = 5 mbar.
– Plateau time Tplat % = 0%.
– Trigger = OFF.
The manufacturer's default settings can be
adjusted in Customer Service Mode. Refer to the
"Set startup settings" section.
Instructions for Use Oxylog 2000 plus SW 1.n