Kikusui PLZ164W User Manual page 31

Plz-4w series
Hide thumbs Also See for PLZ164W:
Table of Contents

Advertisement

Connection procedure of the load input terminal on the rear panel
WARNING
CAUTION
PLZ-4W
• Do not touch the load input terminal while the PLZ-4W is turned ON, as it
may lead to electric shock.
In addition, be sure to use the load input terminal cover.
• The load input terminals on the front and rear panels are coupled inside
the PLZ-4W. A voltage applied to one end will appear on the other end.
• Do not connect equipment to the load input terminals on the front and rear
panels simultaneously, as it can cause damage.
• There is a danger of breakdown. Do not connect the equipment under test
to the load input terminal when the load is turned on.
• There is a danger of overheat. Attach crimping terminal to the wire and use
the set of screws that came with the package for connection.
1.
Turn off the POWER switch.
2.
Check that the output of the equipment under test is off.
3.
Connect the load wire to the load input terminal on the rear panel.
Use the load input terminal cover. See Fig. 2-12 on how to use the cover.
For the connection procedure of the load wire, see Fig. 2-14.
4.
Fix the load input terminal cover to the rear panel using the lock plate.
Pass the inner pin of the lock plate through the hole on the side of the cover.
For details, see Fig. 2-15.
5.
Connect the load wire to the output terminal of the equipment under
test.
6.
Check the polarity of the connection.
Be sure to use the
set of screws
provided with the
package.
Fig. 2-14
Connection to the load input terminal on the rear panel
Bolt (M8 x 18)
Spring washer (M8)
Crimp terminal
Load input terminal
cover
Nut (M8)
Installation and Preparation 2-15