Kikusui PLZ164W User Manual page 175

Plz-4w series
Hide thumbs Also See for PLZ164W:
Table of Contents

Advertisement

Exiting from the calibration screen.
NOTE
PLZ-4W
■ High range calibration
13. Return to number 1 of step E and calibrate the High range (calibration
number "5. CV 150V") by carrying out a similar procedure. Set the volt-
age of the power supply to 155 V or greater.
Step 12 completes the CV mode calibration.
Table 8-5
CV mode settings
Calibration number and item
4
CV 15V
CV 150V
5
1.
Press the PREV (SHIFT+
The original screen displayed before entering calibration appears.
2.
Press the MENU (SHIFT+SET/VSET) key.
The original screen displayed before entering menu setup appears.
• If you press the NEXT (SHIFT+
screen, the calibration data is written to the internal memory.
If you only wish to check the calibration data, be sure to exit from the gain cali-
bration screen by pressing the PREV (SHIFT+
VSET) key.
Output setting of the
power supply
Offset
OffsetVoltage: 15.5 V
Current: 0.3 A
Gain
Offset
Voltage: 155 V
Current: 0.3 A
Gain
) key.
) key or ENTER key on the gain calibration
Maintenance and Calibration 8-13
Voltage to be matched
1.50 V
0.00075 V
15.0 V
0.0075 V
15.0 V
0.0075 V
150 V
0.075 V
) key or MENU (SHIFT+SET/