Kikusui PLZ164W User Manual page 149

Plz-4w series
Hide thumbs Also See for PLZ164W:
Table of Contents

Advertisement

J1-J2 cable
Slave unit
1 unit
2 units
3 units
4 units
CAUTION
NOTE
PLZ-4W
J1
J2
J2
J1
J1
J2
J2
J1
J1
J2
Fig. 6-49
Parallel connection of two slave units
Table 6-13 shows the relationship between the number of slave units and the capacity.
Table 6-13 The number of units connected in parallel and the capacity
PLZ164W
/164WA
66 A / 330 W
99 A / 495 W
132 A / 660 W
165 A / 825 W
• When carrying out parallel operation, be sure to use the load input termi-
nal on the rear panel. Do not connect other equipment to the load input ter-
minal on the front panel.
• Improper connection of the J1 and J2 connectors can damage the PLZ-4W.
• Use the shortest load wire and flat cable possible for the connection. Use a
load wire of sufficient thickness by taking into account the current being used.
It is recommended to use the bus bar instead of wiring the cable.
• Separate the load cable from the flat cable as much as possible to prevent unstable
operation.
+
Master unit
DC INPUT
1000W
1.5
-
150V
0
-
200A
+
Slave unit 1
DC INPUT
1000W
1.5
-
150V
0
-
200A
+
Slave unit 2
DC INPUT
1000W
1.5
-
150V
0
-
200A
Maximum current/Maximum power
PLZ334W
132 A / 660 W
264 A / 1 320 W
198 A / 990 W
396 A / 1 980 W
264 A / 1 320 W
528 A / 2 640 W
330 A / 1 650 W
660 A / 3 300 W
+
Equipment
under test
Load wire
PLZ664WA
PLZ1004W
400 A / 2 000 W
600 A / 3 000 W
800 A / 4 000 W
1 000 A / 5 000 W
Applied Operation 6-63