Calibration number
and item
1
CC(Low)
2
CC(Mid)
3
CC(High)
8-10 Maintenance and Calibration
Step D: Calibration of the gain of the internal reference voltage for
protection function setting
12. Press the ENTER key.
The gain calibration (CC (Low) Limit Gain Adjust) screen appears.
13. Press the
or
knob. Set the current flowing through the shunt resistor within 0.1 % of
the value corresponding to 100 % of the range full scale.
See Table 8-3 for the current settings of each model.
14. Press the ENTER key.
The load is automatically turned off.
The calibration of the low range current is complete.
■ Mid range calibration
15. Return to number 1 of step A and calibrate the mid range (calibration
number "2. CC (Mid)") by carrying out a similar procedure.
■ High range calibration
16. Return to number 1 of step A again and calibrate the High range (cali-
bration number "3. CC (High") by carrying out a similar procedure.
Step 16 completes the CC mode calibration.
Table 8-3
CC mode settings
Output setting of
the power supply
Offset
Gain
Voltage: 5 V
Offset
Current:
Rated current of
Gain
the load device
Offset
Gain
CURSOR key to select DAC LIM, and turn the rotary
Current to be matched
PLZ164W
PLZ334W
PLZ164WA
33 mA
66 mA
0.033 mA
0.066 mA
330 mA
660 mA
0.33 mA
0.66 mA
330 mA
660 mA
0.33 mA
0.66 mA
3.3 A
6.6 A
0.0033 A
0.0066 A
3.3 A
6.6 A
0.0033 A
0.0066 A
33.0 A
66.0 A
0.033 A
0.066 A
PLZ664WA
PLZ1004W
132 mA
200 mA
0.13 mA
0.20 mA
1.32 A
2.00 A
0.013 A
0.020A
1.32 A
2.00 A
0.013 A
0.020A
13.2 A
20.0 A
0.013 A
0.020 A
13.2 A
20.0 A
0.013 A
0.020 A
132.0 A
200.0 A
0.132 A
0.200 A
PLZ-4W