Kikusui PLZ164W User Manual page 170

Plz-4w series
Hide thumbs Also See for PLZ164W:
Table of Contents

Advertisement

CC mode calibration (calibration number 1, 2, and 3)
Procedure
A
B
C
D
A
B
C
D
A
B
C
D
8-8 Maintenance and Calibration
Carry out calibration on the low range items first according to steps A to D. Then,
carry out calibrations for Mid range and High range.
Table 8-2
Overview of the procedure
Percentage with
Current
respect to the full
range
scale (%)
10
Low
100
10
Mid
100
10
High
100
■ Low range calibration
Step A: Calibration of the offsets of internal reference voltage for out-
put setting and measured value
1.
Connect a shunt resistor that matches the value corresponding to 10 %
of the low range full scale.
2.
Press the
or
(Low)".
3.
Connect a CV power supply to the load input terminal and apply 5 V.
Set the current of the power supply approximately in the range of 2 % to
5 % of the rated current of the load device.
4.
Press the ENTER key.
The load automatically turns on, and the offset calibration (CC (Low) Offset
Adjustment) screen appears.
Internal reference
voltage for output set-
ting
Offset
Gain
Offset
CURSOR key to select the calibration number "1. CC
Calibration item
Internal reference
Measured value
voltage for protection
function setting
Gain
Offset
Gain
PLZ-4W