Kikusui PLZ164W User Manual page 173

Plz-4w series
Hide thumbs Also See for PLZ164W:
Table of Contents

Advertisement

CV mode calibration (calibration number 4 and 5)
Procedure
E
F
G
H
E
F
G
H
PLZ-4W
Carry out calibration on the 15-V range items first according to steps E to H. Then,
carry out calibration on the 150-V range items.
The shunt resistor is not used, but you can leave it connected.
Table 8-4
Overview of the procedure
Percentage with
Voltage
respect to the full
range
scale (%)
10
15V
100
10
150V
100
■ Low range calibration
Step E: Calibration of the offsets of internal reference voltage for out-
put setting and measured value
1.
Press the
or
2.
Connect a CC power supply to the load input terminal and supply 0.3 A.
Set the voltage of the power supply to 15.5 V or greater.
3.
Press the ENTER key.
The load automatically turns on, and the offset calibration (CV 15 V Offset
Adjustment) screen appears.
CV 15V
Offset Adjustment
DAC REF : 0x0001
MON : 0x0001
PREV
Fig. 8-8
CV 15 V Offset Adjustment screen
Internal reference
voltage for output
setting
Offset
Gain
CURSOR key to select the calibration number "4. CV 15 V".
Displays the current
calibration value
(hexadecimal) first.
Cursor
NEXT
Indicates hexadecimal.
Calibration item
Internal reference
Measured value
voltage for protection
function setting
Offset
Gain
Offset
Maintenance and Calibration 8-11
Gain