Fault Messages - Dräger X-AM 2500 Technical Manual

Hide thumbs Also See for X-AM 2500:
Table of Contents

Advertisement

Troubleshooting
8.2

Fault messages

Special symbol »
« and
displayed numerical code:
102
103
104
105
106
107
108
109
111
112
113
114
115
116
201
202
203
204
205
207
221
222
223
301
302
303
304
24
Cause
The customer's service life counter has elapsed Reset the service life counter using Dräger
The instrument is defective
Check sum error program code
Bump test interval elapsed
The calibration interval has elapsed (at least
1 calibration interval has elapsed)
Bump test error (at least 1 channel has a bump
test error)
The instrument is defective
The menu function cannot be carried out
because of an error
Faulty alarm element test: Alarm light
Faulty alarm element test: Alarm horn
Faulty alarm element test: Vibration motor
Faulty visual inspection
Instrument deactivated by X-dock
Faulty software update
No valid
zero point calibration of the
DrägerSensor CatEx 125 PR
No
valid
sensitivity
calibration
DrägerSensor CatEx 125 PR
Measurement
from
DrägerSensor
125 PR is in negative range
DrägerSensor CatEx 125 PR not plugged in
or faulty
Error during bump test of DrägerSensor
CatEx 125 PR
Faulty rise time test
Too little oxygen to operate the DrägerSensor
CatEx 125 PR
No valid
zero point calibration of the
DrägerSensor
CatEx
125 PR
conduction
No
valid
sensitivity
calibration
DrägerSensor
CatEx
125 PR
conduction
No valid zero point calibration of the Dräger
Sensor XXS EC1
No valid sensitivity calibration of the Dräger
Sensor XXS EC1
The measured value of DrägerSensor XXS EC
1 is in the negative range
DrägerSensor XXS EC1 is not inserted or faulty Check Dräger Sensor XXS EC1, page 21.
Remedy
CC Vision.
The instrument must be repaired by Service.
The instrument must be repaired by Service.
Run the bump test page 14.
Run sensitivity calibration, see page 18 and/or
page 19.
Run bump test, page 14 or run sensitivity
calibration, page 18 and/or page 19.
The instrument must be repaired by Service.
Determine the error code via the info menu and
switch it off, if necessary.
Repeat alarm element test using X-dock.
Repeat alarm element test using X-dock.
Repeat alarm element test using X-dock.
Repeat visual inspection using X-dock.
Instrument activated using X-dock.
The instrument must be repaired by Service.
Conduct the fresh air calibration page 17.
of
the
Run sensitivity calibration, see page 18 and/or
page 19.
CatEx
Conduct the fresh air calibration page 17.
Check DrägerSensor CatEx 125 PR, page 21
Repeat bump test, where necessary, calibrate or
replace the DrägerSensor CatEx 125 PR, page 21.
Repeat rise time test using X-dock.
Operate sensor in an environment with at least
10 vol. % O
.
2
Conduct the fresh air calibration page 17.
for
heat
of
the
Run
heat
conduction
for
heat
see page 18 and/or page 19.
Conduct the fresh air calibration page 17.
Run sensitivity calibration, see page 19 and/or
fresh air calibration, page 17.
Conduct the fresh air calibration page 17.
sensitivity
calibration,
Dräger X-am 2500