Fault - Cause - Remedy; Reprocessing; Information On Reprocessing - Dräger DigiFlow Manual

Electronic flowmeter
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 18
EnglishEnglish
Display
Action
Basic screen
Function key is
and timer
pressed briefly.
Function key is
not activated.
l/min
timer
min
Function key is
pressed and
held down
(>2 seconds).
Basic screen
Function key is
and "total liter"
pressed briefly.
Function key is
not activated.
l/min
total
l
Reset: Func-
tion key is
pressed and
held down
(>2 seconds).
24
Result
Display
Basic screen
Basic screen
and timer
and "total
hours"
The current val-
ue of the timer
(remaining time
l/min
in minutes until
total
h
signal sounds),
or "OFF", or "- -
-" is displayed.
The timer is
programmed
(5, 10, 15, 30,
45, 60, 90,
120 minutes,
OFF).
Releasing the
Fault – Cause – Remedy
function key
starts the timer
Fault
and counts
down the min-
No display
utes.
Basic screen
and "total liter"
"total liter"
Numerical val-
since last reset
ue 0 L/min, al-
is shown.
though thumb
wheel is open
"total liter" and
"total hours"
are reset to ze-
ro.
The display
shows "over-
flow".
Action
Result
Function key is
Basic screen
pressed briefly.
and "total
hours"
Function key is
"total hours"
not activated.
since last reset
are shown.
Reset: Func-
"total hours"
tion key is
and "total liter"
pressed and
are reset to ze-
held down
ro.
(>2 seconds).
Function key is
Basic screen
pressed briefly.
Cause
Remedy
Battery dis-
Replace bat-
charged.
tery.
Device defec-
Call
tive.
DrägerService.
No oxygen
Restore O
2
available.
supply, check
connection.
Filter in gas in-
Call
let blocked.
DrägerService.
The current
Turn thumb
flow is too high.
wheel clock-
wise to reduce
flow.
Fault
Cause
Function key is
Software error.
deactivated.
The device is
Fall/impact.
visibly dam-
aged, loose
parts inside.
The display
Error on the
shows
PCB.
"Er 001",
"Er 002",
"Er 004",
"Er 008".

Reprocessing

Information on reprocessing

Instructions for reprocessing are based on
internationally accepted guidelines, e.g., standard
ISO 17664.
Instructions for use DigiFlow
Remedy
Remove and
reinsert bat-
tery.
Call
DrägerService.
Call
DrägerService.

Advertisement

Table of Contents
loading

Table of Contents