4.4.29 SENSe:ANALyze:PARameters? Command
Syntax:
SENSe:ANALyze:PARameters?
Parameters:
none
Description:
Query current Analysis parameters.
Example:
sens:anal:par? --- 0.050000, –60.000000, 3.000000,10.000000<END>
Response:
<event loss>,<reflectance>,<end loss>,<pon loss><END>
Errors:
None
4.4.30 SENSe:ANALyze:AUTO Command
Syntax:
SENSe:ANALyze:AUTO<wsp><value>
Parameters:
<value>
Boolean format
Range: 0|1
0 = Auto-analysis is OFF.
1 = Auto-analysis is ON.
Description:
Set ON/OFF analysis to be performed automatically after the test is complete.
Example:
sens:anal:auto 0
Response:
None
Errors:
(–104, "std_wrongParamType, Data Type Error")
4.4.31 SENSe:ANALyze:AUTO? Command
Syntax:
SENSe:ANALyze:AUTO?
Parameters:
None
Description:
Query if auto-analysis after test is set to ON.
Example:
sens:anal:auto? --- 0<END>
Response:
Possible responses are:
0 = Auto-analysis is OFF.
1 = Auto-analysis is ON.
Errors:
None