Chapter 4 OTDR Commands
4.3.4 SOURce:RANge:AVAilable? Command
Syntax:
SOURce:RANge:AVAilable?
Parameters:
None
Description:
Queries the list of available ranges for current wavelength settings.
Example:
sour:ran:ava? --- 5.0, 10.0, 20.0, 50.0, 100.0, 200.0, 300.0<END>
Response:
List of available ranges in kilometers (KM).
Errors:
None
4.3.5 SOURce:RANge Command
Syntax:
SOURce:RANge<wsp><value>
Parameters:
<value>
Numeric format
Range: Model dependent, numeric value returned by available ranges query
command
Description:
Set current range
Example:
sour:ran 100
Response:
None
Errors:
(–200, "std_execGen, Test is Active")
(–224, "std_illegalParmValue, Invalid Parameter Value")
4.3.6 SOURce:RANge? Command
Syntax:
SOURce:RANge?
Parameters:
None
Description:
Query current range.
Example:
sour:ran? --- 50.0<END>
Response:
The value is current range in kilometers (KM).
Errors:
None
4.3.7 SOURce:RESo:AVAilable? Command
Syntax:
SOURce:RESo:AVAilable?
Parameters:
None
Description:
Queries the list of available resolution flags for current range settings.
Example:
sour:res:ava? --- 0, 1, 2<END>
Response:
List of available resolutions: 0-Low Density, 1-High Density, 2-Very High Density
Errors:
None
4-10