Anritsu MT9083 Series Operation Manual page 71

Access master remote control
Table of Contents

Advertisement

4.4.18 SENSe:BCURsor Command
Syntax:
SENSe:BCURsor<wsp><value>
Parameters:
<value>
Floating point format
Range: 0.0 – Current distance range.
Description:
Set B cursor position. Cursor position units – Km.
Example:
sens:bcur 20.5
Response:
None
Errors:
(–224, "std_illegalParmValue, Invalid Parameter Value")
4.4.19 SENSe:BCURsor? Command
Syntax:
SENSe:BCURsor?
Parameters:
None.
Description:
Queries current B cursor position.
Example:
sens:bcur? --- 20.5<END>
Response:
Current B cursor position value in Km.
Errors:
None
4.4.20 SENSe:LSALeft Command
Syntax:
SENSe:LSALeft<wsp><start>,<stop>
Parameters:
<start>
Floating point format
Range: –100.0 – 400.0.
<stop>
Floating point format
Range: –100.0 – 400.0.
Start value must be less or equal to stop value.
Description:
Set start and stop for left LSA marker. Start and stop units – kilometers (KM).
Example:
sens:lsal 0.0, 0.5
Response:
None
Errors:
(–224, "std_illegalParmValue, Invalid Parameter Value")
(–200, "std_execGen, LSA Inactive State")
4.4
SENSe Subsystem Commands
4
4-19

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents