Opc? Operation Complete Query; Rst Reset Command - Anritsu MT9083 Series Operation Manual

Access master remote control
Table of Contents

Advertisement

Chapter 3 Platform SCPI Commands

3.1.7 *OPC? Operation Complete Query

Syntax:
*OPC?
Description:
A device is in the Operation Complete Query Active State (OQAS) after it has
executed *OPC?. The device returns to the Operation Complete Query Idle State
(OQIS) whenever the No Operation Pending flag is TRUE, at the same time placing
a "1" in the Output Queue.
The following actions cancel the *OPC? query (and put the instrument into
Operation Complete, Command Idle State):
. Power-on
Parameters:
None
Response:
1<END>
Example:
*OPC? −> 1<END>

3.1.8 *RST Reset Command

Syntax:
*RST
Description:
The ReSeT command *RST sets the instrument to Top Menu. Pending *OPC
actions are cancelled. The instrument is placed in the idle state awaiting a
command. The *RST command clears the error queue.
The following are not changed:
. Output queue
. Service Request Enable register (SRE)
. Standard Event Status Enable register (ESE)
Parameters:
None
Response:
None
Example:
*RST
3-6

Hide quick links: