Anritsu MT9083 Series Operation Manual page 72

Access master remote control
Table of Contents

Advertisement

Chapter 4 OTDR Commands
4.4.21 SENSe:LSALeft? Command
Syntax:
SENSe:LSALeft?
Parameters:
None
Description:
Query left LSA values.
Example:
sens:lsal? --- 0.0, 0.5<END>
Response:
Start and stop for left LSA marker. Start and stop units – kilometers (KM).
Errors:
(–400, "std_queryGen, LSA Inactive State")
4.4.22 SENSe:LSARight Command
Syntax:
SENSe:LSARight<wsp><start>,<stop>
Parameters:
<start>
Floating point format
Range: –100.0 – 400.0.
<stop>
Floating point format
Range: –100.0 – 400.0.
Start value must be less or equal to stop value.
Description:
Set start and stop for right LSA marker. Start and stop units – kilometers (KM).
Example:
sens:lsar 0.0, 0.5
Response:
None
Errors:
(–224, "std_illegalParmValue, Invalid Parameter Value")
(–200, "std_execGen, LSA Inactive State")
4.4.23 SENSe:LSARight? Command
Syntax:
SENSe:LSARight?
Parameters:
None
Description:
Query right LSA values.
Example:
sens:lsar? --- 0.0,0.5<END>
Response:
Start and stop for right LSA marker. Start and stop units – kilometers (KM).
Errors:
(–400, "std_queryGen, LSA Inactive State")
4-20

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents