Anritsu MT9083 Series Operation Manual page 74

Access master remote control
Table of Contents

Advertisement

Chapter 4 OTDR Commands
4.4.26 SENSe:ORL:MODE Command
Syntax:
SENSe:ORL:MODE<wsp><value>
Parameters:
<value>
Integer format
Range: 0|1|2
0 = A Cursor
1 = Origin
2 = Full Trace
Description:
Set current ORL Mode.
Example:
sens:orl:mode 0
Response:
None
Errors:
(–224, "std_illegalParmValue, Invalid Parameter Value")
4.4.27 SENSe:ORL:MODE? Command
Syntax:
SENSe:ORL:MODE?
Parameters:
None
Description:
Query current ORL Mode.
Example:
sens:orl:mode? --- 0<END>
Response:
Possible responses are:
0 = A Cursor
1 = Origin
2 = Full Trace
Errors:
None
4.4.28 SENSe:ANALyze:PARameters Command
SENSe:ANALyze:PARameters<wsp><spliceloss>,<reflectance>,<endloss>,<pon loss>
Syntax:
Double values:
Parameters:
<splice loss> Range: 0.01 – 9.99
<reflectance> Range: –70.0 - –20.0
<end loss> Range: 1 - 99
<pon loss> Range: 1.0 – 30.0
Set Analysis Parameters for the next test.
Description:
sens:anal:par 0.05,–60.0,3.0,10.0
Example:
None
Response:
(–224, "std_illegalParmValue, Invalid Parameter Value")
Errors:
4-22

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents