Anritsu MT9083 Series Operation Manual page 70

Access master remote control
Table of Contents

Advertisement

Chapter 4 OTDR Commands
4.4.14 SENSe:VOFFset Command
Syntax:
SENSe:VOFFset<wsp><value>
Parameters:
<value>
Floating point format
Range: Offset value can be set plus/minus current dynamic range. (–64.0 to 64.0)
Description:
Set vertical offset for the displayed trace(s). Offset value units – DB.
Example:
sens:voff –5.0
Response:
None
Errors:
(–224, "std_illegalParmValue, Invalid Parameter Value")
4.4.15 SENSe:VOFFset? Command
Syntax:
SENSe:VOFFset?
Parameters:
None
Description:
Queries vertical offset for the displayed trace(s).
Example:
sens:voff? --- 0<END>
Response:
Current vertical offset value.
Errors:
None
4.4.16 SENSe:ACURsor Command
Syntax:
SENSe:ACURsor<wsp><value>
Parameters:
<value>
Floating point format
Range: 0.0 – Current distance range.
Description:
Set A cursor position. Cursor position units – KM.
Example:
sens:acur 20.5
Response:
None
Errors:
(–224, "std_illegalParmValue, Invalid Parameter Value")
4.4.17 SENSe:ACURsor? Command
Syntax:
SENSe:ACURsor?
Parameters:
None.
Description:
Queries current A cursor position.
Example:
sens:acur? --- 20.5<END>
Response:
Current A cursor position value.
Errors:
None
4-18

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents