Controlling Message Sync - Anritsu MT9083 Series Operation Manual

Access master remote control
Table of Contents

Advertisement

Chapter 2 Before use

2.7 Controlling Message Sync

2-14
The following messages (12 types max.) can be received during
measurement by this instrument and analysis of measurement results.
However, if a message is sent to change the measurement parameters
before the previous processing is completed, the message is discarded and
the correct measurement conditions will not be set.
The following program message changes the wavelength to 1550 nm
while performing averaging measurements at a wavelength of 1310 nm.
SOUR:WAV 1310; INIT; SOUR:WAV 1550
Figure 2.7-1 shows the message execution sequence when this message is
sent to the instrument. After setting the initial wavelength, sweeping
starts when INIT is sent. Although a command to change the wavelength
to 1550 nm is sent during sweeping, the command is ignored during
measurement.
SOUR:WAV
1310
Set Wavelength
Figure 2.7-1 Message Execution Sequence
The control for processing the next command after completing processing
of the message sent first is called sync control.
Sync control is performed by the following methods.
 Using *WAI command
 Using *OPC? query
 Using *OPC command and *ESR? query
 By querying execution end
The *WAI command, *OPC? query and *OPC command can be used for all
messages.
This command is ignored.
SOUR:WAV
1550
Change wavelength
INIT: Measuring
Time