Reprocessing Of The Assistant Unit; Preparing The Instrument Holder; Preparing The Connector For The Suction Hose - KaVo Primus 1058 Life Care Instructions

Hide thumbs Also See for Primus 1058 Life:
Table of Contents

Advertisement

Care instructions Primus 1058 Life
2 Setup methods according to DIN EN ISO 17664 | 2.5 Reprocessing of the assistant unit
▪ Microcide AF Liquid (Schülke & Mayr)
▪ FD 322 (Dürr Dental)
▪ Incidin Liquid (Ecolab)
Note
The exterior of the camera and camera hose can be disinfected with chemical
disinfectants.
Comply with the manufacturer's instructions for use with regard to the applic-
ation range. and exposure time.
Material damage due to use of non-approved disinfectant.
Damage to the camera.
▶ Use approved disinfectants only.

2.5 Reprocessing of the assistant unit

2.5.1 Preparing the instrument holder

The instrument holder on the assistant unit is prepared in the same way as the
instrument holder on the dentist unit.
See also:
2 2.4.2 Reprocessing of the instrument holder of the TM table, Page 11

2.5.2 Preparing the connector for the suction hose

▶ Remove connectors ① and ② from the suction hoses.
Cleaning and disinfection of the suction hose connector
▶ Clean the disassembled parts under running water.
▶ Thermodisinfect the disassembled parts for at least 10 minutes at 93 °C
(total cycle: approximately 20 minutes)
Sterilising the suction hose connector
Damage caused by sterilisation.
▶ Do not sterilise the suction hose connector.
Note
Sterilise the suction cannulas according to the manufacturer's instructions.
NOTICE
NOTICE
20 / 44