7 Reprocessing steps in accordance with DIN EN ISO 17664 | 7.3 Packaging

7.2.2 Automated disinfection

The following parts of the unit are released for machine-based disinfection:
▪ Handpiece tray
▪ Motor and motor cable
Note
Please comply with the corresponding Instructions for use when you re-process the
motor and motor cable.
The re-processing steps for the handpiece tray are as follows:
KaVo recommends washer disinfectors in compliance with EN ISO 15883-1 that are
operated with alkaline cleaning agents.
The validations were conducted with the VARIO-TD program, the cleaning agent
neodisher® MediClean and the neutralisation agent neodisher® Z.
▶ For program settings as well as cleansers and disinfectants to be used, please
refer to the Instructions for Use of the thermodisinfector.

7.2.3 Drying

Note
Coolant hose with accessories is intended for single use only and is not to be disin-
fected and sterilised. No drying required.
▶ Allow all disinfected and and sterilised parts to dry fully on room air before using
them again.
Automatic Drying
The drying procedure is usually part of the cleaning programme of the washer disin-
fector.
▶ Follow the instructions for use of the thermodisinfector.

7.2.4 Service, inspection and testing after preparation

Note
It is essential to comply with the hygiene requirements (sterility) during the test after
reprocessing. If sites of fracture and clear changes of the surface are visible, the
parts need to be checked by the Service.
Check for cleanliness, intactness, servicing, and repair as described in the following:
▶ Check the adjustable functions of the unit and the motor function.
▶ Check the hose pump for sufficient coolant flow rate.
▶ Check the control commands on the wireless foot control.

7.3 Packaging

Note
The quality and use of the sterilisation packaging must comply with applicable
standards and be suitable for the sterilisation procedure!
Instructions for use MASTERsurg LUX Wireless REF 1.009.1200
61 / 76