ssembly Replacement Sequence
The following steps show the sequence to replace an assembly in an HP 8753D network
analyzer.
1. Identify the faulty group. Refer to Chapter 4, \Start Troubleshooting Here." Follow up
with the appropriate troubleshooting chapter that identies the faulty assembly.
2. Order a replacement assembly. Refer to Chapter 13, \Replaceable Parts."
3. Replace the faulty assembly and determine what adjustments are necessary. Refer to
Chapter 14, \ ssembly Replacement and Post-Repair Procedures."
4. Perform the necessary adjustments. Refer to Chapter 3, \ djustments and Correction
Constants."
5. Perform the necessary performance tests. Refer to Chapter 2, \System Verication and
Performance Tests."
5-2 Power Supply Troubleshooting