CLEANING THE VENTILATOR

PLANNED MAINTENANCE

Frequency
As necessary
Every 500 hours or 1
month or whichever
comes first
1500 hours or 3
months of use
3,000 hours or every 6
months of use
(whichever comes
first)
Annually
Clean the exterior of the ventilator when necessary with a soft damp cloth
moistened with one of the following cleaning solutions:
Water
Mild detergent or soapy water
Do not allow liquid to penetrate into the ventilator.
Observing the following maintenance intervals helps keep the iVent
operating trouble-free. Follow the iVent
instructions.
Table 13: Planned Maintenance Procedures
Part
Ventilator enclosure
Patient tubing
Cooling air filter and
vents
O2 sensor
Air Inlet Filter
O2 System
Battery
Ventilator
O2 System
Battery
Ventilator
Air Inlet Filter
Inlet muffler
assembly.
Outlet muffler
assembly
201 Service Manual procedure for
TM
Maintenance
Wipe clean the
exterior
Replace and
perform O.V.T.
Clean and replace
air filter as needed
Replace the O2
sensor
Remove the used air
inlet filter and
replace it with a new
inlet filter
O
Calibration
2
Charge the battery
Perform VVT
O2 calibration
Deep discharge and
recharge
Perform VVT
Replace Air Inlet
Filter
Replace Inlet muffler
assembly and Outlet
muffler assembly
174
201
TM