setting same as video.
CH1 to measure a square wave of 10K Hz and CH2 to measure a video signal. The
alternate setting as below:
1、Press "Trig menu"
2、Press "F1" and choose type as "Alternate"
3、Press "F2" and choose source as "CH1"
4、Press "F3" to choose type as "Edge"
5、Press "F4" to choose slope as "Rising"
6、Press "F5" to choose "Next menu"
7、Press "F2" to choose coupling as "DC"
8、Press "F3" till sensitivity display as 0.5 div. CH1 trigger setting finished.
9、Press "F1" to choose "Previous menu"
10、Press "F2" to choose source as "CH2"
11、Press "F3" to choose type as "Video"
12、Press "F4" to choose polarity as "Normal"
13、Press "f5" to choose "Next menu"
14、Press "F2" and choose synchronization as "Line". CH2 trigger setting finished.
Acquired waveform refer to Fig.40

How to Operate the Function Menu

The function menu control zone includes 6 function menu buttons and 3
immediate-execution buttons: SAVE/REL, MEASURE, ACQUIRE, UTILITY,
Fig. 40 Alternate trigger Oscillogram