User's Manual of OWON Color Mixed Signal Digital Storage Oscilloscope
Cursor measurement time menu as below:
Function
Setting
Increment
Time
Frequency
Cursor 1
Time
Cursor 2
Time
Follow up below steps to measure data pulse frequency in CH00 with cursor
measurement:
1. Press "9(CURSOR)" and cursor menu appears
2. Press "F1" till cursor display as "time". Two purples vertical lines display in
panel and refer to Cursor 1 and Cursor 2 separately.
3. Turn "CH1 POSITION" knob and set Cursor 1 position in falling edge of
CH00.
4. Turn "CH2 POSITION " knob and set Cursor 2 position in next falling edge of
CH00
5. Now the time increment display between two cursors is "200us" and frequency
is "5KHz". The time display of Cursor 1 correspond to trigger position is "780us" in
"F3". The time display of Cursor 2 correspond to trigger position is "580us" in "F4".
Instruction
The time difference between two cursors.
The frequency difference between two cursors.
The time display of cursor 1 corresponds to trigger
position.
The time display of cursor 2 corresponds to trigger
position.
Fig 88:Cursor time measurement
88