How to Operate the Function Menu

The function menu control zone includes 4 function menu buttons: Utility,
Measure, Acquire, Cursor, and 2 immediate-execution buttons: Autoset,
Run/Stop.

How to Set the Sampling/Display

Push the Acquire button, the Sampling and Display menu is shown in the right as
follows:
Function
Menu
Acqu Mode
Type
Persist
XY Mode
Counter
Persist
When the Persist function is used, the persistence display effect of the picture
tube oscilloscope can be simulated. The reserved original data is displayed in fade
color and the new data is in bright color.
(1) Push the Acquire button.
4.Advanced User Guidebook
Setting
Description
Sample
Normal sampling mode.
Peak Detect
Use to capture maximal and minimal samples.
Finding highest and lowest points over
adjacent intervals. It is used for the detection of
the jamming burr and the possibility of reducing
the confusion.
Average
It is used to reduce the random and don't-care
noises, with the optional number of averages.
Turn the M knob to select 4, 16, 64, 128 in the
left menu.
Dots
Only the sampling points are displayed.
Vect
The space between the adjacent sampling
points in the display is filled with the vector
form.
OFF
1 Second
2 Seconds
Set the persistence time
5 Seconds
Infinity
ON
Turn on/off XY display function
OFF
ON
Turn on/off counter
OFF
27