Cursor Measurement - Owon MSO5022 User Manual

Portable mixed signal digital storage oscilloscope
Table of Contents

Advertisement

User's Manual of OWON Color Mixed Signal Digital Storage Oscilloscope
The description of the cursor measurement menu is shown as the following table:
Function Menu Setting
Type
Source
Delta
Cursor 1
Cursor 2

Cursor Measurement

When carrying out the cursor measurement, the position of Cursor 1 can be adjusted
with the CURSOR1 (VERTICAL POSITION) knob of Channel 1, and that of
Cursor 2 can be adjusted with the CURSOR2 (VERTICAL POSITION) knob of
Channel 2.
Perform the following operation steps for the voltage cursor measurement of the
channel CH1:
1. Press CURSOR and recall the CURS MEAS menu.
2. Press the F1 menu selection button to choose Voltage for Type, with two purple
dotted lines displayed along the horizontal direction of the screen, which
indicating CURSOR1 and CURSOR2.
3. Press the F2 menu selection button and choose CH1 for Source.
4. Adjust the positions of CURSOR1 and CURSOR2 according to the measured
waveform, with the absolute value of the voltage amplitude difference between
Cursor 1 and Cursor 2 displayed in the increment window. The current position
of Cursor1 is displayed in the Cursor1 window and that of Cursor2 is displayed in
the Cursor2 window (see Fig. 58).
Description
OFF
Switch off the cursor measurement.
Voltage
Display the voltage measurement cursor and
menu.
Time
Display the time measurement cursor and menu.
CH1, CH2
Choose the channel generating the waveform to
which the cursor measurement will be applied.
Read the difference between cursors.
Read the position of Cursor 1 (the Time is read
with reference to the horizontal trigger position
and the Voltage is to the ground point).
Read the position of Cursor
with reference to the horizontal trigger position
and the Voltage is to the ground point).
61
2
(the Time is read